ZGS-S series internal water cooling generator DC generator
Scope of application
It is suitable for the DC leakage current and DC withstand voltage test of the stator winding in the water-cooled generator state
Performance characteristics
(1) The output current is large, which can meet the leakage current and DC withstand voltage test of the internal water-cooled generator under water;
(2) Separate structure design of doubler barrel and intermediate frequency variable, light weight and portable;
(3) Electronic control and tuning are adopted for easy operation. Voltage, current, etc. are directly read on the panel meter without conversion;
(4) Timing and timing reminder function;
(5) Perfect over-voltage and over-current protection function, and external zinc oxide over-voltage protection.
Technical indicators: can be customized according to customer requirements
Output voltage (kV)Output power (W)
specification Technical Parameters | 60/80 | 60/120 | 60/200 | 80/200 | 80/300 | 90/150 |
60 | 60 | 60 | 80 | 80 | 90 | |
Output current (mA) | 80 | 120 | 200 | 200 | 300 | 150 |
4800 | 7200 | 12000 | 16000 | 24000 | 13500 | |
Voltage measurement error | 1.0% (full scale) ±1 word | |||||
Total current measurement error | 1.0% (full scale) ±1 word | |||||
Leakage current measurement error | ±2.0% | |||||
Overvoltage setting error | ≤1% | |||||
Voltage stability | 2.0% (random fluctuation, power supply voltage change ±10%) | |||||
Way of working | Rated load, continuous work ≤10 minutes at a time | |||||
Ambient temperature | -10℃~40℃ | |||||
相Relative humidity | Relative humidity <85% non-condensing | |||||
Altitude | Below 2000 meters |